Electrophysiology: general
Principle
Electrophysiology is the study of the electrical properties of biological cells
and tissues. It involves measurements of voltage change or electrical current
flow by electrodes in various systems, from single ion channel proteins to single
neurons (particularly action potentials) and whole tissues like the heart.
Classical electrophysiological techniques
The principal types of electrodes are:
1) metal discs (singles or arrays),
2) metal tracings on printed circuit boards.
2) needles (singles or arrays), diameter μm
scale, inserted into or just outside a single cell.
4) small glass pipettes, also for extra- or
intracellular recording.
The principal preparations include:
· living organisms,
· excised tissue (acute or cultured),
· dissociated cells from excised tissue (acute or
cultured),
· artificially grown cells or tissues,
· hybrids of the above.
The pipette techniques has various versions:
· With a low impedance pipette an extracellular placement
may pick up the activity of several nearby cells simultaneously, and this is
termed multi-unit recording (also with a metal microelectrode).
· With the electrode tip (tip size ca. 1 μm,
resistance some MΩ) more closely to the neuron, the recording is termed
single unit recording (also with metal microelectrodes, resistance some 100
kΩ).
· With the pipette tip pressed against the cell membrane,
to which it tightly adheres. This is a semi-intracellular recording with spikes
of some 5-10 mV. (pipette resistance a few tens of MΩ).
· The electrolyte within the pipette may be brought into
fluid continuity with the cytoplasm by delivering a pulse of pressure to the
electrolyte in order to rupture the small patch of membrane encircled by the pipette
rim (intracellular or whole cell recording, resistance some 50 MΩ). The
electrolyte may contain a drug. When the pipette contains a tracer (e.g. HRP)
then the resistance can be about 200 MΩ.
· Alternatively, ionic continuity may be established by
"perforating" the patch by allowing exogenous ion channels within the
electrolyte to insert themselves into the patch (perforated patch recording).
· The patch may be left intact (patch recording).
· Finally, the patch is so small that it comprises only one
channel: a single channel recording.
As electrode size increases, the resolving
power decreases. Larger electrodes are sensitive only to the net activity of
many cells, the so called local field potentials (see More Info). Still larger electrodes, such as non insulated needles
and surface electrodes (discs) used by clinical (EEG, and surgical
neurophysiologists, are sensitive only to certain types of synchronous activity
within populations of cells numbering in the millions.
Optical electrophysiological techniques
Optical electrophysiological techniques were
created to overcome the limitations that electrical activity is recorded at approximately
a single point within a volume of tissue. Interest in the spatial distribution
of bioelectric activity prompted development of molecules capable of emitting
light in response to their electrical or chemical environment. Examples are voltage
sensitive dyes and fluorescent proteins.
With one or more such compounds administrated
via perfusion, injection or gene expression, the distribution of electrical
activity may be observed and recorded.
Application
Many particular electrophysiological techniques
and recordings have abundant clinical application:
Electrocardiography (ECG), for the heart;
Electroencephalography (EEG), for the brain and Electrocorticography for the
cerebral cortex;
Electromyography (EMG) for the muscles
Electro-oculography,
for the eyes
Electroretinography,for
the retina
Electro-olfactography
for the olfactory receptors;
Evoked
potentials for auditory, visual, somatosensory assessment.
Bioelectric Recognition Assay (BERA)
BERA is a novel method for measuring changes in the membrane
potential Em of cells immobilized in a gel matrix. Apart from the
increased stability of the electrode-cell interface, immobilization preserves
the viability and physiological functions of the cells. BERA is primary used in
biosensor applications in order to assay analytes which can change Em
in a characteristic, ‘signature-like’ way. BERA has been used for the detection
for human viruses, veterinary disease agents and plants in a highly specific,
rapid (1-2 minutes), reproducible and cost-efficient fashion. The method has
also been used for the detection of environmental toxins, such as herbicides
and the determination of very low concentrations of superoxide anion in
clinical samples. A recent development of the BERA technology is the technique
called Molecular Identification through Membrane Engineering (MIME). This technique allows for
building cells with absolutely defined specificity against virtually any molecule
of interest, by embedding thousand of artificial receptors into the cell
membrane.
More info
Extracellular recording
A microelectrode (tip size ca. 1 μm), into the brain, nerve or sensory
tissue (e.g. retina) of a living animal will usually detect the activity of one
neuron or axon. The spikes of a "single unit" recording are very like
the intracellular spikes, but they are much smaller (typically 0.1-1 mV).
Multi-unit recordings are often used in
conscious animals to record changes in the activity of a discrete brain area
during some behavior of the animal. Recordings from one or more such electrodes
which are closely spaced can be used to identify the number of cells around it
as well as which of the spikes come from which cell. This process is called
spike sorting and is suitable in areas where there are identified types of
cells with well defined spike characteristics.
If the electrode tip is bigger still,
generally the activity of individual neurons cannot be distinguished but the
electrode will still be able to record a field potential generated by the
activity of many cells. Such an electrode inserted in or near the auditory
nerve can record a compound action potential.
Intracellular recording
Intracellular
recording involves measuring voltage and/or current
across the membrane of a cell with (generally) a glass micropipette (< 1μm
and a resistance of tens of MΩ) filled with an electrolyte and connected
with a metal wire (generally an Ag wire coated with AgCl), to make a connection
between the electrolyte and the amplifier. The coating reduces the galvanic
potential between electrolyte and silver wire. This is necessary since the
galvanic potential is very large (+1.98 V). Moreover, it reduces electrode
noise which is higher the smaller the contact surface.
The pipette must be inserted inside the cell,
so that the membrane potential can be measured (in rest −60 to −80
mV). Micropipettes are filled with a solution that has a similar ionic
composition to the intracellular fluid of the cell. The voltage measured by the
electrode is compared to the voltage of a reference electrode, usually an
Ag-AgCl wire in contact with the extracellular fluid somewhere in the tissue.
Pipette impedance
In general, the smaller the electrode tip,
the higher its electrical resistance so an electrode is a compromise between
size (small enough to penetrate with minimum damage) and resistance (low enough
so that small neuronal signals can be discerned from thermal noise in the
electrode tip, see below). Pipettes have also a capacitance, as does the electrode
cable and the amplifier has an input capacitance. With a capacitance poor cable
(also short, and with a driven guard, i.e. the shield has the amplifier input
voltage by using a buffer amplifier) and capacitance poor amplifier, these parallel
capacitances together can be brought down to about 10 pF.
The pipette restistance Rpip is:
Rpipette = Rshaft + Rtip + Rmedium,
Rpipette = relectrolyteL shaft/(¼πdshaft2)
+ (2/dtip─ 2/dshaft)relectrolyteLtip/(½πdshaft)
+ rmedium/(2πdtip), (1)
where d is diameter and L is length. With relectrolyte
= 75 Ωcm, = rmedium
= 750 Ωcm, L shaft =
Rpipette = 75 kΩ + 11.9 MΩ + 2.4 MΩ.
The far majority of Rpipette is
due to the tip geometry. With dtip fixed, Rpipette can only be reduced by reducing Ltip.
With an electrode resistance 100 MΩ the resulting low pass first order
filter (see Linear
first order system) has a cut
off frequency of 160 Hz (time constant 1 ms), hardly enough to record intracellular
spikes. Since the pipette capacitance is determined by the thickness (reciprocally)
and the surface (linear) of the glass wall (the dielectricum), the geometry of
the tip is of great importance.
Sharp electrode technique
For recording the potential inside the cell
membrane with minimal effect on the ionic constitution of the intracellular
fluid a sharp electrode can be used. They look like patch clamp pipettes but the
pore is much smaller so that there is very little ion exchange between the intracellular
fluid and the electrolyte in the pipette. The resistance of the electrode is
tens to hundreds of MΩ. Often the tip of the electrode is filled with
various kinds of voltage sensitive dyes like Lucifer yellow to be injected by Electrophoresis. A positive or negative, DC or pulsed voltage is applied
to the electrodes depending on the polarity of the dye. Later, this enables
identification of the cell under a microscope.
Field potentials

Fig. 1
A schematic diagram of a field potential recording from a rat
hippocampus. When the synapse releases glutamate the postsynaptic membrane
opens ionotropic glutamate receptor channels. The net flow of current is
inward, so a current sink is generated. A nearby electrode (#2) detects this as
negativity. An intracellular electrode placed inside the cell body (#1)
records the change in membrane potential that the incoming current causes.
Extracellular
field potential
Extracellular field potentials are local current sinks or sources that are generated by the collective
activity of many cells. Usually a field potential is generated by the
simultaneous activation of many neurons by synaptic transmission. Fig. 1 to the
right shows a hippocampal synaptic field potential. The lower trace shows a
negative wave that corresponds to a current sink caused by positive charges
entering cells through postsynaptic glutamate receptors, while the upper trace
shows a positive wave that is generated by the current that leaves the cell (at
the cell body) to complete the circuit.
Local
field potential
A local field potential (LFP) is recorded using a low
impedance extracellular microelectrode, placed sufficiently far from individual
local neurons to prevent any particular cell from dominating the signal. The
unfiltered signal reflects the sum of action potentials from cells within
approximately 50-350 μm from the tip of the electrode and slower ionic
events from within 0.5-
The amplifier measures the electrical potential
difference between the microelectrode and a reference electrode,
placed somewhere else in the body or tissue with a similar extracellular medium
(to cancel the both electrochemical polarization effects at the interface of
both metal electrodes (or the Ag/AgCl wires in the pipette).
Synchronized Input
The LPF is believed to represent the synchronized input
into the observed area. The quick fluctuations, caused by the short inward and
outward currents of the action potential, are filtered out, leaving only the
slower fluctuations. Therefore the spike plays no part in the LFP, which thus
comprises the more sustained currents in the tissue, typical of somato-dendritic
origin. The major slow current is the PSP. It was thought until recently that
EPSP’s and IPSP’s were the exclusive constituents of LFP’s. However, phenomena
unrelated to synaptic events have been found to contribute to the LFP.
Geometrical Arrangement
Cells which contribute to the slow field variations are
determined by the geometric configuration of the cells themselves (as with the
pyramidal cells.)
When there is simultaneous activation of the dendrites a
strong dipole is produced. His may even give rise to a signal recordable with
EEG. In cells where the dendrites are arranged more radially, in a plane, the
potentials between individual dendrites and the soma tend to cancel.
Thermal
noise
The thermal noise generated in a conductor (e.g.a resistor),
Vn (amplitude, measured as the effective (rms) value, in Volt) is:
Vn =
2(kTR∙Δf)0.5, (2)
where k is Boltzmann’s constant, T the absolute temperature
(K), R the resistance (Ω), and Δf the bandwidth of the amplifier.
For extracellular recording Vn hardly plays a
role since R is low. As example: R = 2.5 MΩ, T = 300 K, and Δf is 300
Hz, then Vn = 3.5 μV, whereas spikes are generally larger than 10
μV. For intracellular recordings the same holds. With R = 250 MΩ, Vn
= 35 μV, just small enough to find the spikes extracellular before
impaling the cell. However, in addition there are other noise sources: amplifier
noise; with a digital output the noise in the analog-to-digital converter; noise
generated at the interface of the metal electrode (or the Ag/AgCl wires in the
pipette); noise generated by the opening of an ion channel caused by the net
flow of ions into the cell from the extracellular medium, or out of the cell
into the extracellular medium. Which type of noise
dominates depends on the precise conditions of the recording.