Echography
Principle
Echography, also called medical (ultra)sonography
is an ultrasound-based
diagnostic imaging technique.
Echography uses a probe containing acoustic transducers
(generally of piezo crystals, see piezoelectricity) to send sound into
a material (here tissue). Whenever a sound wave encounters a material with a
different acoustical
impedance, part of the sound wave is reflected (see Waves)
which the probe detects as an echo,
see Fig. 1.

Fig. 1
The time it takes for the echo to travel back to the
probe is measured and used to calculate the depth of the tissue interface
causing the echo. The greater the difference between acoustic impedences, the
larger the echo is. The reflectioncoefficient R is:
(1)
with Ar and Ai the
amplitudes of reflected and impinging wave and Z1 and Z2
the acoustic impedences of medium 1 and 2 respectively. When these media are
respectively water (Z=152.000 rayl) and air (Z=400 rayl) then R=0.999474, which
is equivalent to a transmission loss of 20log(1-R)=65.6 dB. When medium 1 is
air and 2 is water, the same holds (see (1)). Consequently: with a large ratio of the both
impedances, the reflection is large.
The above consideration does not take in account scatter
from the object, which diminishes the reflectance and disturbes imaging. Taking
water as substitute for blood, R of blood-muscle interface is only 0.034 (see
for values Acoustic
impedance), which asks for highly sofisticated hardware and
software to obtain a good image (noise reduction). When bone is involved, R is
some 2-20 times higher.
A water-based gel ensures good acoustic coupling between
skin and the ultrasound scan head.
A 2D- image can be obtained by a probe with many
transducers and a 3-D images can be constructed with a specialized probe.
From the amount of energy in each echo, the difference in
acoustic impedance can be calculated and a colour is then assigned accordingly.
Limitations
The spatial resolution in the axial direction (the depth)
is directly related to the wavelength λ of a pure ultrasound frequency. In the lateral
direction the resolution is determined by the width (the aperture angle) of the
beam due to divergence.
Further, there is an ambiguity in depth position. This
occurs when the time lapse between sending and receiving a wave is larger than
the period time tper, The reflected wave of the objects at all
distances n times 0.5λ are superimposed at the reflected pulse of the
object itself. Mathematically: depth = (0.5trecieve/tper,
− integer{0.5trecieve/tper, −})c + nλ. This problem is solved by sending short pulses and adjusting the pulse
interval time such that any reflection from boundaries to at least the depth of
interest are arriving within the pulse interval time. Since a pulse comprises
many frequencies (see Signal Analysis and Fourier) the received signal needs some complicated computation (deconvolution)
to reconstruct the echo-image.
The echographic modes
In the A-mode the strength, i.e. amplitude, of the reflected wave is indicated on the
vertical axis and time at the horizontal one, with time zero at the origin.
In the B-mode the strength is indicated by the brightness (grayscale) of a dot. With
the B-scan, along the vertical axis the penetration depth is indicated. The
beam of the ultrasound changes slightly its angle of incidence every time a new sound pulse is
emitted. In this way a kind of section of the anatomical object is obtained.
However, the less depth, so the closer to the sound source, the more compressed
is the image in the horizontal direction (so parallel to the surface). Such a
scan is made many times a second and in this way a kind of movie is made. This
is very helpful for moving structures, especially the heart (Fig. 2).

Fig. 2 Abnormal echocardiogram showing a mid-muscular
ventricular septum defect in a new-born
child. The red line in the ECG mark the time instant that the image was
captured. Colors indicate blood velocity measured with the combined Doppler
apparatus.
In the M-mode, movement is visualized by displaying
time along the horizontal axis and an image is made for a single beam
direction. When for instance this beam impinges on a mitralis valve, the image
shows opening and closing of the valve (Fig. 3).

Fig. 3 Echocardiogram in M-mode
Technical strengths
Ø It images soft tissues very well and is particularly useful for delineating
the interfaces between solid and fluid-filled spaces.
Ø It renders "live" images.
Ø It shows the structure as well as functional aspects.
Ø Widely available and comparatively flexible.
Ø Small, cheap, easily carried scanners (bedside) available.
Ø Classical ultrasound devices have trouble penetrating bone but current
research on ultrasound bone imaging will make it possible with dedicated
devices in the future.
Ø Performs very poorly when there is a gas between the scan head and the
organ of interest, due to the extreme differences in acoustical impedance.
Ø Even in the absence of bone or air, the depth penetration of ultrasound is
limited, making it difficult to image structures that are far removed from the
body surface, especially in obese patients.
Ø The method is operator-dependent. A high level of skill and experience is
needed.
Echography is widely utilized, often with a hand-held
probe. It is especially practiced in cardiology, gynecology and obstetrics,
urology (kidney, prostate), vascular medicine, gastroenterology (also liver),
endocrinology and ophthalmology.
More
info
There exist several types of echograpy, mostly combined
with a Doppler application.
Echography can be enhanced with Doppler measurements,
which employ the Doppler effect (see Doppler principle) to assess
whether structures (usually blood) are moving towards or away from the probe,
and its relative velocity. By calculating the frequency shift of a particular
sample volume, for example a jet of blood flow over a heart valve, its velocity
and direction can be determined and visualized. This is particularly useful in
cardiovascular studies and vascular examinations of other organs (e.g. liver
portal system). The Doppler signal is often presented audibly using stereo
speakers: this produces a very distinctive, although synthetic, sound. Doppler
echography can by distinguished in several modifications. The most common ones
are here discussed.
The duplex (Doppler) scanner
The duplex scanner detects in a selected
part of the image the moving blood by using the Doppler effect. The scanner
calculates the actual velocity of the blood provided the angle between the direction of the ultrasonic
beam and the direction of movement is known. The operator therefore aligns a
marker along the direction of flow in the blood vessel and positions a cursor
at the height of the peak systolic blood velocity.
In the common and superficial femoral
arteries, the waveform normally has a forward component followed by a reverse
component and a second smaller forward component. This is called a triphasic
waveform because of the three phases. More distally in the superficial femoral
artery, the second forward component may be absent, giving a biphasic waveform
with two phases.

Fig. 4 Duplex
scan of superficial femoral artery with a triphasic blood velocity waveform Horizontal axis: time (large ticks 1 s);
vertical axis velocity.
The frequency shift is normally in the audio range
(due to the ”ultra” frequencies), so
most duplex scanners send the signal to a pair of audio speakers, and this
enables the operator to hear the signal in addition to seeing the display (as in Doppler echocardiography).
Color
Doppler scanners
Color Doppler (CD) scanners detect and display moving
structures by superimposing color onto the grey-scale image. Color is
superimposed wherever the scanner detects a moving structure, usually blood.
The hue of the color shows the direction and magnitude of the blood velocity.
In this image, red and yellow indicate flow away from the probe, with dark red
representing low velocities and orange and yellow indicating higher velocities.
Flow towards the probe is indicated in blue and green, with green indicating
higher velocities. The hue can therefore be used to identify sites where the
artery becomes narrower and the blood has to move faster to achieve the same
volume flow rate. When the blood velocity exceeds the limit of the color scale,
aliasing occurs (high velocity in one direction interprets as lower velocity in
the other, wrong, direction). Color Doppler can also be used to display venous
blood flow.
CD and Duplex sonography are often combined to
Duplex/CD sonography, especially for assessing stenoses.
Power
Doppler (PD)
Duplex/CD sonography is not an effective technique
when the artery under study is almost perpendicular to the ultrasonic beam or by other poor
conditions as bowel gas, breathing movements and obesity. Power Doppler (PD)
has improved diagnostic capabilities of vascular Duplex/CD sonography, mainly
because it is independent from the angle of insonation and has more sensitivity.
PD generates an intravascular color map reflecting the integrated power in the
Doppler signal, which essentially depends on the density of red blood cells
within the sample volume. However, due to its intrinsic limitations, PD cannot
replace conventional sonographic techniques and especially CD. So, PD is used
as an adjunctive tool in vascular sonography.
Tissue Doppler Imaging (TDI)
Tissue Doppler Imaging (TDI) measures and displays peak velocities high
temporal resolution (ca. 8 ms) at any point of the ventricular wall during the
cardiac cycle. The mean velocities can be calculated with time velocity
maps and displayed as color encoded velocity maps in either an M-mode or a
two-dimensional format. Indeed, since all the points within the ventricular
walls are velocity-encoded in real-time, the color-coded display should provide
a huge amount of information which could form the basis for the application of
accurate, reproducible quantitative evaluation.
http://en.wikipedia.org/wiki/Ultrasonography
http://www.worldwidewounds.com/2000/sept/Michael-Lunt/Doppler-Imaging.html#Colour%20Doppler%20scanners