CT scan (dual
energy)
Principle
.
Fig. 1 CT
apparatus
Computed tomography (CT),
originally known as computed axial
tomography (CAT or CT scan)
and body section röntgenography (see X-ray machine),
is a medical imaging method (techniques and processes used to create images of
parts of the human body for clinical purposes). It employs tomography (imaging
by sections) where digital geometry processing (design of algorithms for
3D-modeling) is used to generate a 3D image of the internals of an object from
a large series of 2D X-ray (electromagnetic radiation with 10-0.01 nm
wavelengths, see also Spectroscopy)
scans. The data of CT scans taken around a single axis of rotation can be manipulated
by a process known as windowing
(see Imaging: windowing). Standard CT
generates images in the axial or transverse plane which can be reformatted in
various planes or as volumetric (3D) representations of structures.
Fig. 2 Principle
of spiral CT
The principle of operation is an X-ray source that
rotates around the object with X-ray sensors positioned on the opposite side of
the circle from the X-ray source. X-ray slice scans are taken one after the
other as the object is gradually passed through the gantry. The data of the
slices are combined together by the mathematical procedure known as tomographic
reconstruction, for instance to generate 3D volumetric information (3D-CT
scan), which are viewable from multiple different perspectives on CT
workstation monitors. This can also be done with helical or spiral CT
machines, which generate high-resolution 3D images from the data of the moving
individual slices.
Application
Although most common in medicine, CT is also used in
other fields, for example in experimental earth sciences, material sciences and
archeology.
CT has become an important tool in medical imaging to
supplement X-rays and ultrasonography or echography. Although it is still
expensive, it is the gold standard in diagnostic. It is also used in preventive
medicine or screening. Some important fields of application are described
below.
Cardiology
Abdominal and
pelvic region Cancer, acute
abdominal pain, organ injury, disorders resulting in morphological changes of
internal organs, often with contrast (barium sulfate for fluoroscopy, (see also Fluorescence), iodinated
contrast for pelvic fractures).
Further CT is useful for (complex) fractures in the extremities.
CT is inadequate for osteoporosis (radiation doses,
costs) compared to DXA scanning for assessing
bone mineral density (BMD), which is used to indicate bone strength.
More Info
Fig.3 ..CT scanner with cover removed to show the
principle of operation
Working
principle
In conventional CT machines, an X-ray tube (see X-ray machine)and detector are physically rotated
behind a circular shroud (see Fig. 3). In electron beam tomography (EBT) the tube is far larger and higher power to support the high
temporal resolution. The electron beam is deflected in a hollow funnel shaped
vacuum chamber. X-rays are generated when the beam hits the stationary target.
The detector is also stationary. Contrast materials are used to highlight
structures such as blood vessels and to obtain functional information.
The data stream, representing the varying radiographic
intensity sensed, reaches the detectors on the opposite side of the circle during
each sweep. Then, it is computer-processed to calculate cross-sectional
estimations of the radiographic density, expressed in Hounsfield
units (HU).
Hounsfield unit
This unit is defined as:
HUmaterial
= 1000(μmaterial -
μwater)/μwater,
where μ is the attenuation coefficient, which is
analogue to the parameter A (absorbance or extinction) of the Lambert-Beer law.
HU is dependent of the beam intensity of the X-ray source (mostly given in
kilo-electron volt, keV).
Since μwater is by definition zero, the radiodensity of distilled water at STPD
conditions (see Gas volume units, STPD,
BTPS and ATPS) is defined as 0 HU. At 80 keV, that of air at STPD is ca. -1000 HU, cancellous bone amounts to 400 HU
and cranial bone to 2000 HU. The attenuation of metallic implants (dental,
extremities) depends on the element’s atomic number. Titanium has ca. 9000 HU
(at 80 keV) and iron steel ca. 24500, which can completely extinguish the X-ray
and is therefore responsible for well-known line-artifacts in computed tomograms.
Pixels in an image obtained by CT scanning are displayed
in terms of HUs (from -1024 to +3071). When the CT slice thickness is also
factored in, the volumetric unit is known as a voxel, which is a cubical pixel.
The phenomenon that one part of the detector cannot differ between different
tissues is called the Partial Volume
Effect. That means that a big amount of cartilage and a thin layer of
compact bone can cause the same attenuation in a voxel as hyperdense cartilage
alone.
Dual source (or energy) scanning offers
the potential of differentiating materials beyond the visualization of
morphology – for example, direct subtraction of either vessels or bone during
scanning. Dual
Advantages
over Projection Radiography (see
Radiography)
·
CT completely eliminates the superposition of
images of structures outside the area of interest.
·
Because of the inherent high-contrast resolution of
CT, differences between tissues that differ in physical density by less than 1%
can be distinguished.
·
Data from a single CT imaging procedure consisting
of either multiple contiguous or one helical scan can be viewed as images in
the axial, coronal, or sagittal planes. This is referred to as multiplanar
reformatted imaging.
CT is regarded as a moderate to high radiation diagnostic
technique. Unfortunately the newer CT technology requires higher doses for better
resolution. For instance, CT angiography avoids the invasive insertion of an arterial
catheter and guide wire and CT colonography may be as good as barium enema for
detection of tumors in the large intestines, but at the cost of a higher dose.
Cardiac MSCT is equivalent of 500 chest X-rays in terms
of radiation. The risk on breast cancer is not well established. The positive
predictive value is approximately 82% while the negative prediction is ca. 93%.
Sensitivity is ca. 81% and the specificity is about 94%. The real benefit in
the test is the high negative predictive value.
The radiation dose for a particular study depends on
multiple factors: volume scanned, patient build, number and type of scan
sequences, and desired resolution and image quality.
|
Examination |
Typical
effective dose (mSv*) |
|
Chest
X-ray |
0.02 |
|
Head
CT |
1.5 |
|
Abdomen |
5.3 |
|
Chest |
5.8 |
|
Chest,
Abdomen, Pelvis |
9.9 |
|
Cardiac
CT angiogram |
6.7-13 |
|
CT
colongraphy |
3.6
- 8.8 |
*The sievert (Sv) has the same dimensions as the gray
(i.e. 1 Sv = 1 J/kg = 1 m2·s–2), but the former measures
the biological effect and the later the radiated dose.
Adverse reactions to contrast agents
Because CT scans often rely on IV-contrast agents, there
is a low but non-negligible level of risk associated with the contrast agents
themselves, like (life-threatening) allergic reactions to the contrast dye
(e.g. causing kidney damage).